Temperature range change
Please add this feature to cart if you want to change the temperature working range of the controller.
Applicable only for the products that have this item in their accessory list.
Product Description
If you want to buy TEC controller with the changed temperature range, add this service (item) to your cart and specify new temperature range in customer message during checkout.
Attention! Changing the stabilization temperature range may change stabilization accuracy. This is caused by working on different voltage levels inside the device.
In the table below you can see temperature stabilization accuracy values for different temperature ranges. You can see that actual accuracy depends on range spread not on the exact temperature when the range starts for example.
Please refer to the following abbreviations:
Tmin – minimum temperature
Tmax – maximum temperature
Tstab min – minimum temperaturę stabilization accuracy (at a higher temperature of the range)
Tstab max – maximum temperaturę stabilization accuracy (at a lower temperature of the range)
Temperature range = Tmax - Tmin
Tmin [⁰C] |
Tmax [⁰C] |
Tstab max [⁰C] |
Tstab min [⁰C] |
-30 |
0 |
0,05 |
0,1 |
0 |
30 |
0,06 |
0,1 |
30 |
60 |
0,07 |
0,1 |
60 |
90 |
0,08 |
0,11 |
Tmin [⁰C] |
Tmax [⁰C] |
Tstab max [⁰C] |
Tstab min [⁰C] |
-30 |
30 |
0,05 |
0,34 |
30 |
90 |
0,07 |
0,2 |
If you need high temperature stabilization accuracy, temperature range is much more important than what Tmin and Tmax you choose:
Temperature range spread |
Tstab max [⁰C] |
Tstab min [⁰C] |
10 |
0,05 |
0,06 |
20 |
0,05 |
0,07 |
30 |
0,05 |
0,11 |
40 |
0,05 |
0,12 |
50 |
0,05 |
0,17 |
60 |
0,05 |
0,34 |